Model:
|
S-2120
|
|
Light
Source:
|
Tungsten-halogen
lamp & deuterium lamp
|
|
Monochromator:
|
Modified
Czerney-Turner type with 1200 lines/mm holographic grating
|
|
Detector:
|
UV
enhanced silicon photodiode
|
|
Wavelength
Range:
|
190
~
1100nm
|
|
Wavelength
Accuracy:
|
<±0.5nm
|
|
Wavelength
Reproducibility:
|
<
±
0.1nm
|
|
Spectral
Band Pass Width:
|
<0.1nm
|
|
Photometric
Range:
|
-3.0
~
3.0 A, 0 to 125 %T
|
|
Wavelength
Slew Rate:
|
5000nm/min
|
|
Photometric
Accuracy:
|
<1%
at 1 A
|
|
Photometric
Reproducibility:
|
<0.001
A
|
|
Noise
Level:
|
<0.001A
|
|
Stability:
|
±0.002
A/hr
|
|
Drift:
|
0.002
A/hr at 340nm
|
|
Stray
Light:
|
0.05%T
at 220 &340 nm
|
|
Display
Mode:
|
Backlight,
graphical, liquid crystal display with contrast control
|
|
Power
Requirements:
|
220±10%,
110V, 50-60Hz
|
|
Physical
Dimensions:
|
541mm
(W) x 430mm (D) x 243mm (H)
|
|
Weight:
|
16kg
|
|
Operating
Temperature:
|
5°C
~
40°C
|
|
Humidity:
|
Less
than 95% relative, non-condensing
|