Model:
|
S-3130
(S-3150)
|
|
Light
Source:
|
Pre-aligned
Tungsten & Deuterium Lamp
|
|
Detector:
|
1024
channels PDA
|
|
Wavelength
Range:
|
190
– 850nm (190 – 1150nm)
|
|
Wavelength
Accuracy:
|
±
0.4nm (0.5nm) with Holmium oxide glass filter
±0.4nm
(0.5nm) with D2 peaks 486.0nm and 656.1nm
|
|
Wavelength
Reproducibility:
|
±
0.02nm
|
|
Wavelength
Resolution:
|
<0.75nm
(<0.95nm)
|
|
Photometric
Range:
|
-3.0
~ +3.0 Au
|
|
Data
Acquisition Mode:
|
Absorbance,
%T, %R, Intensity, Concentration
|
|
Photometric
Accuracy:
|
±
0.005 Au at 440 nm, 590 nm with SSCS F3 filter
±
0.01 Au at 440 nm, 590 nm with SSCS F4 filter
|
|
Photometric
Noise:
|
<0.0002
Au, with integration time in full
|
|
Photometric
Reproducibility:
|
<0.002
Au with 60 ppm Potassium Dichromate
<0.001 Au at 440, 590nm with 1.0 Au of neutral density glass filter
<0.002 Au/hr at zero Au, 500nm
|
|
Photometric
Resolution:
|
<
0.0001 Au
|
|
Photometric
Stability:
|
±
0.003 Au/hr at zero Au, 500 nm
|
|
Stray
Light:
|
<0.03%
at 340nm (NaNO2, ASTM)
<0.05% at 220nm (KCl, EP, ASTM)
<1% at 200nm (KCl, EP, ASTM)
|
|
Baseline
Flatness:
|
<0.0005
Au
|
|
Shortest
Scan Time:
|
0.02
sec for full wavelength range (In Ultra Kinetic Range)
|
|
Typical
Scan Time:
|
1~1.5
sec for full wavelength range
|
|
Integration
Time:
|
0.02
sec ~
6 sec
|