Features of the 6400 Visible and 6405 UV / Visible Spectrophotometers for
a wide range of applications
|
Photometrics |
|
Select
this mode for the routine measurement of:
-
%
Transmittance (0 – 199.9%)
-
Absorbance
(-0.300 to 3.000A)
-
Concentration
(range –300 to 9999)
-
Concentration
by Factor (range 0.000 to 9999.99)
-
Unit
selection (ppm, mg/L, g/L, %, M and blank)
|
| Spectrum |
|
The
6400 & 6405 have full scanning capability with
-
3
scan speeds (166, 400 and 1200 nm/minute)
-
3
scan intervals (0.2, 1 and 5nm)
-
Baseline
correction
-
Instant
display of spectrum
-
Automatic
axis scaling
-
Peak
and valley data display
-
Hard
copy of data/scan with printer accessory
-
2
scans stored in memory
|
|
Multi-wavelength
analysis |
|
For applications
such as DNA/RNA composition @260/280nm, where difference and ratio
analysis can be used quantitatively.
|
|
Kinetics |
|
 |
Software
for Kinetics is included as standard. Temperature control is achieved with
the Peltier cell accessory.
-
Lag
time selection (0 – 9999 seconds)
-
Measurement
time (10 – 9999 seconds)
-
Graphical
display of data
-
Line
of best fit between two user set points
-
Kinetics
factor input and rate calculation
-
Statistical
display gives initial absorbance, mean rate of absorbance change,
kinetic activity value, and equation for line of best fit
|
|
Quantitation
|
|
This
mode is only available when the 8 position Multi-Cell Changer is
fitted
-
Concentration
calculation at a single wavelength from Absorbance or Transmittance
readings
-
Concentration
range, 0 – 9999, auto-ranging to a maximum of three decimal places
-
Automatic
calibration sequence initiated via set-up menu
-
Up
to 7 calibration points can be used
-
Choice
of curve fitting methods:
-Linear
interpolation
-Linear
regression (line of best fit)
-Line
of best fit through zero
-
Print
out of curve and numbered sample readings
-
2
calibration curves stored in memory
|
|
GLP
|
|
Protocols
are supported, by the following standard features:
-
User
can set instrument number (up to 6 digits)
-
Time
and date of last calibration
-
Print
out states sample number, the reading and units of measure, wavelength
of analysis, plus time and date
-
Stored
scans along with concentration curves are stamped according to time
and date.
|
|
Data
Handling / Interfacing |
|
Bi-directional
RS232 Interface to allow remote operation,
calibration and interrogation by an external PC.
-
Fixed
baud rate at 1200 bits/second
-
Results
can be downloaded by a single touch of the print key
-
Analog
output (±
2000mV) is provided for use with a chart recorder
-
Output
is proportional to the displayed reading
(1mV per
0.1%T, 0.001A or 1 concentration unit)
|

|
 |